首页 | 本学科首页   官方微博 | 高级检索  
     检索      

AN AUTOMATED ADIABATIC CALORIMETER FOR USE FROM 80 TO 600 K——HEAT CAPACITIES OF α-ALUMINA
引用本文:张志英.AN AUTOMATED ADIABATIC CALORIMETER FOR USE FROM 80 TO 600 K——HEAT CAPACITIES OF α-ALUMINA[J].中国科学B辑(英文版),1986(12).
作者姓名:张志英
作者单位:Thermochemistry Laboratory,Northwestern University,Xi'an
摘    要:The construction of an automated adiabatic calorimeter for use from 80 to 600 K is reported. The main feature of the calorimoter is to mount three radiation shields in the cryostat for advancing the working tomperature. The molar heat capacity of α-Al_2O_3 has been measured in order to assess the reliability of the calorimeter. Deviations of the experimental heat capacities of α-Al_2O_3 from the smoothed curve are within ±0.3% except for a few experimental points. Our smoothed results agree with those of tho National Bureau of Standards. USA, within ±0.3% over the entire working temperature range.

本文献已被 CNKI 等数据库收录!