首页 | 本学科首页   官方微博 | 高级检索  
     


Reversed-phase high-performance liquid chromatographic separation and electrochemical detection of retinol and its isomers
Authors:W A MacCrehan  E Sch?nberger
Abstract:Baseline separation of the isomers of retinol using reversed-phase high-performance liquid chromatography (HPLC) in less than 30 min is presented. A new approach to the detection of retinol using electrochemical detection is developed. The oxidative electrochemistry of retinol is studied at a glassy-carbon electrode using coulometry, ultraviolet-visible spectrophotometry and HPLC. Amperometric detection in HPLC for retinol provided a linear response from 0 to 1.5 micrograms/ml and a detection limit of 4.1 ng/ml. Electrochemical detection was compared to ultraviolet-visible absorbance detection for the determination of retinol in human serum extracts. Good agreement is found for the results obtained with the two detectors.