首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Interpreting the difference between conventional and bi-directional plate-height measurements in liquid chromatography
Authors:Hui Guo  Douglas D Frey
Institution:Department of Chemical and Biochemical Engineering, University of Maryland Baltimore County, 1000 Hilltop Circle, Baltimore, MD 21250, USA
Abstract:An experimental and theoretical study was conducted of the column characterization technique in which plate heights determined using the conventional pulse-response method are compared with those determined using a bi-directional method where an eluite sample is introduced into one end of a chromatographic column and elution occurs at the same end after the flow direction is reversed inside the column. Experiments are presented for a micropellicular HPLC column before and after its performance has been degraded by repeated sample injections, for a low-pressure column containing nonporous glass particles, and for an HPLC column containing particles with 300 Å pores. The results obtained are interpreted in terms of several different theories which apply in various Fourier number ranges. It was shown that the transcolumn contribution to convective dispersion in a chromatographic column is largely responsible for the difference observed between conventional and bi-directional plate-height measurements and that a collocation method can be employed to develop a useful analytical expression for this contribution.
Keywords:Bi-directional plate-height measurement  Theory of chromatography  Convective dispersion  A term  van Deemter equation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏