Abstract: | Solid-phase extraction with two-step elution has been developed for effective elimination of copper and iron interference with mercury determination by flow-injection cold vapour atomic absorption spectrometry (CV AAS). Sodium tetrahydroborate(III) was used as reducing agent. Cation-exchanger Dowex 50Wx4 was applied for the sorption of mercury and both interfering ions. In the first step elution of Cu(II) and Fe(III) was performed using 0.5 mol L–1 KF solution. Then mercury was eluted with 0.1% thiourea in 8% HCl. The detection limit (3) for Hg(II) was 27 ng L–1. The expanded uncertainty estimated for the whole procedure was about 6%. The accuracy of the proposed method was evaluated by determination of the recovery of known amount of mercury added to mineral, spring, and tap waters, and by analysis of a certified reference material BCR-144R (sewage sludge). |