首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Simultaneous determination of hypoxanthine,inosine, inosine-5′-phosphate and adenosine-5′-phosphate with a multielectrode enzyme sensor
Authors:Etsuo Watanabe  Shunsuke Tokimatsu  Kenzo Toyama  Isao Karube  Hideaki Matsuoka  Shuichi Suzuki
Institution:Department of Food Engineering and Technology, Tokyo University of Fisheries, 4-5-7 Konan, Minato-ku, Tokyo 108 Japan;Research Laboratory of Resource Utilization, Tokyo Institute of Technology, Nagatsuta-cho, Midori-ku, Yokohama 227 Japan
Abstract:A multielectrode enzyme sensor for the simultaneous determination of adenosine-5′-phosphate (AMP), inosine-5′-phosphate (IMP), inosine (HXR) and hypoxanthine (HX)in fish meat was developed by assembling four enzyme sensors for AMP, IMP, HXR and HX in a flow cell. These compounds were determined from oxygen consumption according to the following reactions: AMP AD IMP NT HXR NP, PO3?4 HX XO, O2 Uric acid where AD is AMP deaminase, NT is 5′-nucleotidase, NP is nucleoside phosphorylase and XO is xanthine oxidase. Enzymes were covalently bound to a membrane prepared from cellulose triacetate, 1,8-diamino-4-aminomethyloctane and glutaraldehyde. Sensors for HX, HXR, IMP and AMP were prepared by attaching membranes of XO, XONP, XO NPNT, and of XONPNT and AD, respectively, to four oxygen electrodes. Samples extracted from sea bass, bream, flounder, abalone and arkshell were analyzed within 5 min, from the simultaneous response curves of the four electrodes. Results obtained by the multisensor system were in good agreement with those determined by each single electrode.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏