首页 | 本学科首页   官方微博 | 高级检索  
     


The direct determination of cadmium in urine by electrochemical atomic absorption spectrometry with the L'vov platform
Authors:J.J McAughey  N.J Smith
Affiliation:Health and Safety Executive, Occupational Medicine and Hygiene Laboratory, 403 Edgware Road, London NW2 6LN Great Britain
Abstract:A simple, direct procedure for the measurement of cadmium in urine is described. Graphite-furnace atomic absorption spectrometry is used in conjunction with selective atomisation at 800°C from a L'vov platform. Urine samples are diluted with an equal volume of deionised water and 20-μl aliquots are injected. Calibration is done by standard additions. The sensitivity is 0.016 μg Cd l?1 for 1% absorption for a 20-μl sample. Within-run precision is 4.9% at 0.84 μg l?1. The detection limit is 0.06 μg l?1, which allows normal unexposed levels of cadmium in urine to be determined. The method is applicable to the determination of urinary cadmium levels of both occupationally non-exposed and exposed populations.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号