Rapid determination of sulfide in waste waters by continuous flow analysis and gas diffusion and a potentiometric detector |
| |
Authors: | K Brunt |
| |
Institution: | Analytical Department, Potato Processing Research Institute TNO, 27 Rouaanstraat, 9723 CC Groningen The Netherlands |
| |
Abstract: | The continuous-flow system is based on a flow-through gas diffusion unit and a potentiometric detection with sulfide-selective electrode. The sampling rate is about 15 h?1. In the 0.5–600 mg l?1 sulfide range, the mean slope of the response is 30.2 ± 0.3 mV per decade. The percentage transference of H2S across the teflon membrane is 48 ± 1%. Up to sample concentrations of 50 mg l?1 sulfide, results agree well with those obtained by a distillation method. Sample storage is discussed. |
| |