首页 | 本学科首页   官方微博 | 高级检索  
     


Minimization of matrix interferences in the determination of aluminium silicon by electrothermal atomic absorption spectrometry with the L'vov platform
Authors:M. Taddia
Affiliation:“G. Ciamician” Chemical Institute of the University, 1-40126 Bologna Italy
Abstract:The pyrolytic L'vov platform in a pyrolytically-coated graphite tube was successful in eliminating the suppressive interference by silicon (? 10 mg ml?1) on the aluminium signal, thus allowing ? 0.2 μg Al g?1 to be determined without preliminary separation.
Keywords: