Affiliation: | aDepartamento de Química-CCET, Av Filinto Muller 1555, Caixa Postal 549, CEP 79074-460, Campo Grande, Mato Grosso do Sul, Brazil;bDepartamento de Química, Faculdade de Filosofia, Ciências e Letras de Ribeirão Preto/USP, Av. Bandeirantes, 3900, CEP 14040-901, Ribeirão Preto, São Paulo, Brazil |
| |
Abstract: | This work presents an electroanalytical methodology developed for square-wave voltammetry based in the electrochemical reduction in hanging mercury drop electrode (HMDE), which is simple, fast, reliable and sensitive for determination of moxifloxacin (MOXI) in tablets and spiked urine human samples. The support electrolyte that provided a more defined and intense peak current for MOXI determination was the phosphate buffer 0.04 mol l− 1 pH 8.0. In the best-optimized conditions the drug presented an only peak of reduction at − 1.38 V vs. Ag/AgCl, using an Eacc. of − 0.30 V. An LOD of 0.44 and 3.20 ng ml− 1 and an LOQ of 1.46 and 10.60 ng ml− 1 were found for the pure standard of moxifloxacin and in the presence of matrix, respectively. A good recovery was obtained for assay spiked urine samples and a good quantification of MOXI was achieved in a commercial formulation. The methodology proposed was more sensitive than the spectrofluorimetric and spectrophotometric method with precision and accuracy equivalent. |
| |