Diffuse surface calibration method to improve accuracy and dynamic range of aerosol elastic light scattering measurements
Authors:B M Crosland  M R Johnson  K A Thomson
Institution:1. Energy and Emissions Research Lab., Mechanical and Aerospace Engineering, Carleton University, 1125 Colonel By Drive, Ottawa, ON, K1S 5B6, Canada
2. Measurement Science and Standards, National Research Council, 1200 Montreal Road, Ottawa, ON, K1A 0R6, Canada
Abstract:A new method to calibrate detectors for elastic light scattering (ELS) measurement based on diffuse scattering from a Lambertian surface is presented. The method produces a calibration signal that is approximately seven orders of magnitude larger than a propane gas Rayleigh scattering calibration. The method also allows for calibration of detectors such as photodiodes, which are not sensitive enough to detect Rayleigh scattering for calibration but possess characteristics desirable for the measurement of soot ELS. Since the method is only suitable for backward scattering calibrations, transfer of calibration data from a backward- to a forward-oriented detector is accomplished with a secondary laser and integrating sphere. In demonstration experiments, calibration constants for photomultiplier tube (PMT) detectors obtained using both Rayleigh scattering and diffuse surface scattering agreed within experimental uncertainties as did measurements of in-flame scattering coefficients obtained with PMTs and photodiodes. However, achievable uncertainties with the diffuse-surface calibration approach were significantly reduced. More importantly, by enabling the use of photodiode detectors in ELS measurements, the new method facilitates operation at higher photon fluxes resulting in improved signal-to-noise ratios, reduced influence of photon shot noise, and the ability to achieve higher dynamic range in transient measurements.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号