Magnesium-selective test strip |
| |
Authors: | Capitán-Vallvey L F Ramos M D Fernández Gálvez P Alvarez de Cienfuegos Gómez J |
| |
Affiliation: | Department of Analytical Chemistry, Faculty of Sciences, University of Granada, Avda. Fuentenueva s/n, Granada 18071, Spain |
| |
Abstract: | A selective test strip based on an ion-exchange mechanism to determine magnesium is described. This optical and reversible test strip, made on a transparent polyester sheet, has a circular polymeric film of plasticised PVC that contains all of the reagents necessary to produce a response to magnesium, namely, a cation-selective neutral ionophore, 2-amino-N,N′-diheptyl-N,N′-dimethyl-butanediamide, a chromoionophore, N,N-diethyl-5-(octadecanoylimino)-5H-benzo[a]phenoxazine-9-amine, and potassium tetrakis (4-chlorophenyl)borate (TCPB) as a lipophilic salt, which is evaluated by absorbance measurement at 655 nm in a standard photometer. The experimental variables that influence test strip response, especially in terms of selectivity and response time, have been studied. The test strip responded linearly to magnesium between 0.009 and 160 mM in activities. The reproducibility intermembrane at a medium level of the range was 8.0%, as R.S.D., of log aMg2+, and 5.9% as R.S.D. intramembrane. The procedure was applied to the determination of magnesium in different types of waters (tap, well and mineral) validating the results against a reference procedure. This proposed method is quick, inexpensive, selective and sensitive and uses only conventional instrumentation. |
| |
Keywords: | Magnesium determination Test strip Ionophore-chromoionophore system Water analysis |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|