A Simple Method to Measure Plasma Levels of Propafenone with Fluorescence Detection
Authors:C Flores-Pérez  H Juárez-Olguín  J Flores-Pérez  B Ramírez-Mendiola  J Bobadilla Chávez
Institution:(1) Laboratorio de Farmacología, Instituto Nacional de Pediatría (INP), Avenida Imán No.1 3er piso, Colonia Cuicuilco, CP 04530 México D.F., México;(2) Departamento de Farmacología, Facultad de Medicina, Universidad Nacional Autónoma de México, México;(3) Servicio de Cardiología, INP, México D.F., México
Abstract:The aim of the present study was to develop a simple method to measure plasma levels of propafenone by liquid chromatography with a C18 reverse-phase column and fluorescence detection, without previous derivatization of the sample. Linearity was assessed in the range from 50 to 1000 ng mL−1 and had a correlation coefficient of 0.999. The inter- and intra-day coefficients of variation were below 5%. The limits of detection and quantification were 15 ng mL−1 and 50 ng mL−1, respectively. Drug levels were determined satisfactorily in two patients. A simple and reliable method was developed, especially useful in children with heart failure under treatment with propafenone.
Keywords:Column liquid chromatography  Propafenone  Plasma levels  Pharmacokinetics