HILIC-MS-MS for the Quantification of Pidotimod in Human Plasma
Authors:H S Chen  M Shen  L Y Chen
Institution:1. State Key Laboratory of Organic Geochemistry, Guangzhou Institute of Geochemistry, Chinese Academy of Science, Guangzhou, 510640, China
2. Hygiene Detection Center, School of Public Health and Tropical Medicine, Southern Medical University, Guangzhou, 510515, China
Abstract:A method for fast, sensitive, and specific hydrophilic interaction chromatography combined with tandem mass spectrometry (HILIC-MS/MS) was developed for the first time to determine the level of pidotimod in human plasma. With rosiglitazone as internal standard, analysis was carried out on a HILIC column (150 mm × 2.1 mm, 3.5 ??m) using a mobile phase consisting of methanol:0.2% formic acid (60:40, v/v). Detection was carried out by tandem mass spectrometry using electrospray ionization (ESI). Linear calibration curves were obtained in the concentration range of 11.2?C1.12 × 104 ng mL?1 for pidotimod, with a lower limit of quantification of 11.2 ng mL?1. The intra- and inter-day precision values were high, with standard deviations lower than 15%, and the accuracy, in terms of relative error, ranged from ?10.5 to 9.4% at all quality control (QC) levels.
Keywords:
本文献已被 SpringerLink 等数据库收录!