首页 | 本学科首页   官方微博 | 高级检索  
     


LC Determination of Thiols Derivatized with 4-(Aminosulfonyl)-7-fluoro-2,1,3-benzoxadiazole after SPE
Authors:Ke-Jing Huang  Cong-Hui Han  Jing Li  Zhi-Wei Wu  Yan-Ming Liu  Ying-Ying Wu
Affiliation:1. College of Chemistry and Chemical Engineering, Xinyang Normal University, Xinyang, 464000, China
Abstract:Solid-phase extraction (SPE) coupled with high-performance liquid chromatography?Cfluorescence detection (LC?CFL) was developed for the determination of three thiol compounds including glutathione, cysteine and acetylcysteine. 4-(Aminosulfonyl)-7-fluoro-2,1,3-benzoxadiazole was used for derivatization of thiols. Factors affecting derivatization and extraction efficiency were optimized. Sample solution (2?mL) was extracted on a SPE column for 2?min and then eluted with 400???L methanol. The analytes were injected onto the LC system for separation on a C18 column, and eluted with methanol?Cacetate buffer. The analytes were detected by fluorescence at an emission wavelength of 515?nm with excitation at 385?nm. The linearity of the method was in the range of 0.1?C60???M, with correlation coefficients ranging from 0.9979 to 0.9990. The detection limits of the method were in the range of 5?C20?nM. The proposed method was applied to the analysis of human plasma samples with recoveries of 86?C112.9%.
Keywords:
本文献已被 SpringerLink 等数据库收录!