首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of Inorganic Anions by Capillary IC Using Hexadimethrine Bromide-Modified Silica Stationary Phases
Authors:Li Rong  Zhiguo Liu  Menglin Ma  Lee Wah Lim  Toyohide Takeuchi
Institution:1. Department of Chemistry School of Physics and Chemistry, Xihua University, 999 Jinzhou-Road, Chengdu, 610039, China
2. Department of Chemistry Faculty of Engineering, Gifu University, 1-1 Yanagido, Gifu, 501-1193, Japan
Abstract:A simple, rapid and highly sensitive capillary ion chromatographic method for direct determination of iodide and thiocyanate is reported. Separation was achieved on a laboratory-made capillary column (100 mm × 0.32 mm i.d.) packed with silica gel, followed by modification with 20 mM hexadimethrine bromide. Sodium perchlorate?Cmethanol (95:5, v/v) was used as the eluent and analyte anions were detected at 225 nm. Iodate, bromate, nitrate, iodide and thiocyanate were eluted within 8 min, with relative standard deviations of the retention time, peak area and peak height smaller than 2.4%. Effects of the eluent composition on the retention behavior were also investigated. The limit of detection (S/N = 3) of iodide was 6.5 ??g L?1, whereas that of thiocyanate was 16.2 ??g L?1. The method was successfully applied to the rapid and direct determination of iodide in powdered milk and thiocyanate in human saliva without any pre-concentration. The modified column could be used for about 1 month (8 h operation per day) without loss of hexadimethrine bromide.
Keywords:
本文献已被 SpringerLink 等数据库收录!