首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Comparison of IFCC and NGSP methods for determination of glycated haemoglobin using advanced regression techniques
Authors:Viera Mrázová  Ján Mocák  Adriána Bednárová  Ján Balla
Institution:(1) Chair, IFCC Working Group on HbA1c Standardisation, Isala Klinieken, Location Weezenlanden, Groot Wezenland 20, 8011 JW Zwolle, The Netherlands
Abstract:The objective of this work was a correct statistical comparison of two assay methods for determination of glycated haemoglobin HbA1c. The immunoturbidimetric determination of HbA1c was performed in two ways: using an automatic analyser Hitachi 912, calibrated according to the IFCC reference system (International Federation of Clinical Chemistry and Laboratory Medicine) and using an analyser Advia 1200 and the NGSP reference system (National Glycohaemoglobin Standardization Program). For statistical comparison of these two analytical methods several advanced regression methods were used, which respect random errors of both compared methods. Specifically, Deming regression with and without weights, orthogonal regression, and Passing-Bablok regression were employed. The results demonstrate that the investigated analytical assay methods do not correspond to each other. The summarized results indicate usefulness of better harmonisation of two existing reference systems.