A post-column extraction system for the determination of tertiary amines by liquid chromatography with chemiluminescence detection
Authors:P. J. M. Kwakman  U. A. Th. Brinkman  R. W. Frei  G. J. de Jong  F. J. Spruit  N. G. F. M. Lammers  J. H. M. van den Berg
Affiliation:(1) Department of Analytical Chemistry, Free University, De Boelelaan 1083, 1081 HV Amsterdam, The Netherlands;(2) Duphar B.V., P. O. Box 2, 1380 AA Weesp, The Netherlands
Abstract:Summary The combination of an ion-pair extraction detection system with peroxyoxalate chemiluminescence detection has been investigated. Ion-pairs of protonated tertiary amines with a chemiluminescent counter ion are on-line post-column extracted to 1,2-dichloroethane containing bis(2,4,6-trichlorophenyl)oxalate (TCPO). Hydrogen peroxide is added to the organic phase by means of a solid-state perhydrit reactor. The influence of the base catalyst (imidazole) on the chemiluminescence reaction in apolar solvents was studied. Some sulfonated chemiluminophores were compared with respect to their chemiluminescence, ion-pairing and extraction properties, and 5-dimethylaminonaphthalene-1-sulfonate was found to be the most suitable reagent. The detection limit of the potential drug secoverine is in the sub-nanogram range.
Keywords:Column liquid chromatography  Ion-pair extraction detection  Peroxyoxalate chemiluminescence  Chemiluminescent counter ion