Assay of free captopril in human plasma as monobromobimane derivative,using RPLC/(+)ESI/MS/MS: validation aspects and bioequivalence evaluation |
| |
Authors: | Andrei Medvedovici Florin Albu Iuliana Daniela Sora Stefan Udrescu Toma Galaon Victor David |
| |
Institution: | 1. Labormed Pharma S.A., Splaiul Independentei no. 319 E, Bucharest 060044, Romania;2. University of Bucharest, Faculty of Chemistry, Department of Analytical Chemistry, Sos. Panduri, no. 90, Bucharest, 050663, Romania |
| |
Abstract: | A sensitive method for determination of free captopril as monobromobimane derivative in plasma samples is discussed. The internal standard (IS) was 5‐methoxy‐1H‐benzimidazole‐2‐thiol. Derivatization with monobromobimane immediately after blood collection and plasma preparation prevents oxidation of captopril to the corresponding disulfide compound and enhances the ionization yield. Consequently, derivatization enhances sample stability and detection sensitivity. Addition of the internal standard was made immediately after plasma preparation. The internal standard was also derivatized by monobromobimane, as it contains a thiol functional group. Preparation of plasma samples containing captopril and IS derivatives was based upon protein precipitation through addition of acetonitrile, in a volumetric ratio 1:2. The reversed‐phase liquid chromatographic separation was achieved on a rapid resolution cartridge Zorbax SB‐C18, monitored through positive electrospray ionization and tandem MS detection using the multiple‐reaction monitoring mode. Transitions were 408–362 amu for the captopril derivative and 371–260 amu for the internal standard derivative. The kinetics of captopril oxidation to the corresponding disulfide compound in plasma matrix was also studied using the proposed method. A linear log–log calibration was obtained over the concentration interval 2.5–750 ng/mL. A low limit of quantitation in the 2.5 ng/mL range was obtained. The analytical method was fully validated and successfully applied in a three‐way, three‐period, single‐dose (50 mg), block‐randomized bioequivalence study for two pharmaceutical formulations (captopril LPH 25 and 50 mg) against the comparator Capoten 50 mg. Copyright © 2009 John Wiley & Sons, Ltd. |
| |
Keywords: | captopril monobromobimane derivatization RPLC separation positive electrospray ionization oxidation kinetics nonlinear calibration validation bioequivalence |
|
|