首页 | 本学科首页   官方微博 | 高级检索  
     检索      


On-line coupling of packed column supercritical fluid chromatography with multiple wavelength detector and radioactivity flow-through detector
Authors:P Lembke  H Engelhardt  R Ecker
Institution:(1) Institut für Angewandte Physikalische Chemie, Universität des Saarlandes, 66041 Saarbrücken, Germany;(2) Hewlett Packard, 76337 Waldbronn, Germany
Abstract:Summary The coupling of packed column supercritical fluid chromatography with a commercial ldquoRadioactivity Flow-Through Detectorrdquo for the sensitive and highly selective detection of radioactive nuclides (e.g.14C is described. The radioactivity flow-through detector showed no baseline shift when pressure or modifier (methanol) gradients were applied. The detector cell was pressure resistant even at high flow rates, over a period of approx. 100 hours and showed no leakage problems even at 40°C, 340 bar, 4.0 ml/min and a modifier content up to 20%.
Keywords:Supercritical fluid chromatography  Packed column SFC  Radioactivity flow-through detector  Multiple wavelength detector  Modifier gradient