首页 | 本学科首页   官方微博 | 高级检索  
     


Selective ionization for elemental analysis with an ion trap mass spectrometer
Authors:A. W. Garrett   P. H. Hemberger  N. S. Nogar
Affiliation:

Chemical Science and Technology Division, MS J565, Los Alamos National Laboratory, Los Alamos, NM 87545, USA

Abstract:The ion storage capacity (<106) of ion trap mass spectrometers (ITMS) can sometimes limit the ability to analyze trace components in complex mixtures. We demonstrate here that resonant laser ablation (RLA) can offer a degree of selectivity in the ionization process, thus allowing the preferential accumulation of analyte ions in the trap. Selectivities of 75 and 50, for chromium and iron, respectively, are reported here for RLA of stainless steel in an ITMS. We offer suggestions to improve both the selectivity and the ionization efficiency, relative to the results reported here.
Keywords:Ionization efficiency   Ion trap mass spectrometer   Resonant laser ablation   Selectivity
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏