首页 | 本学科首页   官方微博 | 高级检索  
     


Flow-injection chemiluminescence determination of formaldehyde with a bromate-rhodamine 6G system.
Authors:Baoxin Li  Meilin Liu  Zhujun Zhang  Chunli Xu
Affiliation:Department of Chemistry, School of Chemistry and Materials Science, Shaanxi Normal University, Xi'an, 710062, PR China.
Abstract:In this paper, a novel flow-injection chemiluminescence (CL) system for the determination of formaldehyde is described. It is based on a strong enhance effect of formaldehyde on the weak CL emission of the reaction between potassium bromate and rhodamine 6G in a sulfuric acid medium. A possible mechanism for this CL reaction is proposed. A CL calibration graph was linear in the range of 0.8 - 200 microg l(-1) and the detection limit was 0.3 microg l(-1) (3sigma). The relative standard deviation was less than 3% for 10 microg l(-1) formaldehyde (n = 11). The method has been applied to determine formaldehyde in the air samples.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏