Automated high-performance liquid chromatographic determination of amphetamine in biological fluids using column-switching and on-column derivatization |
| |
Authors: | R. Herráez-Hernández P. Campíns-Falcó S. Díaz-Oltra |
| |
Affiliation: | (1) Departamento de Química Analítica, Facultad de Química, Universidad de Valencia, Doctor Moliner 50, 46100 Burjassot, Valencia, Spain |
| |
Abstract: | Summary A rapid and simple liquid-chromatographic method has been developed for on-line quantification of amphetamine in biological fluids. Untreated samples (20 μL) are injected directly into the chromatographic system and purified on a 20 mm×2.1 mm i.d. pre-column packed with 30 μm Hypersil C18 stationary phase. After clean-up the analyte is transferred to the analytical column (125 mm×4 mm i.d., 5 μm LiChrospher 100 RP18) for derivatization and separation using a mixture of acetonitrile and the derivatization reagent (o-phthaldialdehyde andN-acetyl-L-cysteine) as the mobile phase. The experimental conditions for on-line derivatization and resolution of the amphetamine have been optimized, and the results have been compared with those obtained by derivatizing the analyte in pre-column mode. The method described has been applied to the determination of amphetamine in plasma and urine. Good linearity and reproducibility were obtained in the 0.1–10.0 μg mL−1 concentration range, and limits of detection were 25 ng mL−1 and 10 ng mL−1 with UV and fluorescence detection, respectively. The procedure described is very simple and rapid, because no off-line manipulation of the sample is required; the total analysis time is approximately 8 min. |
| |
Keywords: | Column liquid chromatography Column switching On-column derivatization Amphetamine |
本文献已被 SpringerLink 等数据库收录! |
|