首页 | 本学科首页   官方微博 | 高级检索  
     


Proposed performance qualification and calibration method for evaporative light scattering detectors
Authors:J. L. Sims
Affiliation:(1) Glaxo SmithKline Beecham Pharmaceuticals Research and Development, New Frontiers Science Park, CM19 5AW Harlow, Essex, UK
Abstract:Summary A method for the performance testing of Sedex 55 Evaporative Light Scattering Detectors (ELSD) has been developed as part of our regulatory compliance programme. The method uses a readily available chromatographic column, caffeine as the test probe and can be completed in less than one hour.
Keywords:Column liquid chromatography  Light scattering detection  Performance qualification
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏