首页 | 本学科首页   官方微博 | 高级检索  
     检索      


High-throughput rheology in a microfluidic device
Authors:Schultz Kelly M  Furst Eric M
Institution:Department of Chemical Engineering and Center for Molecular and Engineering Thermodynamics, University of Delaware, Newark, Delaware 19716, USA.
Abstract:High-throughput rheological measurements in a microfluidic device are demonstrated. A series of microrheology samples are generated as droplets in an immiscible spacer fluid using a microfluidic T-junction. The compositions of the sample droplets are continuously varied over a wide range. Rheology measurements are made in each droplet using multiple particle tracking microrheology. We review critical design and operating parameters, including the droplet size, flow rates and rapid fabrication methods. Validation experiments are performed by measuring the solution viscosity of glycerine and the biopolymer heparin as a function of concentration. Overall, the combination of microrheology with microfluidics maximizes the number of rheological measurements while simultaneously minimizing the sample preparation time and amount of material, and should be particularly suited to the characterization of scarce or expensive materials.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号