首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of low concentrations by spectrophotometry: Uranium-thiocyanate system
Authors:A Madhusudan and R K Iyer
Institution:(1) Fuel Reprocessing Division, Bhabha Atomic Research Centre, Trombay, 400085 Bombay, India;(2) Analytical Chemistry Division, Bhabha Atomic Research Centre, Trombay, 400085 Bombay, India
Abstract:Errors in the determination of low concentrations by spectrophotometry are investigated with the uranium-thiocyanate system as an example. The reagent blank has significant absorption and measurements are made at 375 nm instead of the λmax. The error in the intercept of the calibration curve is an important factor in such measurements and the errors involved in the estimation of 1 μg/ml (normal working range 4–40μg/ml) are studied. It is shown that both random and systematic errors associated with the intercept are responsible for observed errors. The two types of errors are resolved by ANOVA (analysis of variance). The error in the measurement of a single value is estimated and compared with measured values for different calibration ranges. It is seen that two factors predominantly influence the error in the measured concentration — the variance from regression and the closeness of the measured value to the mean of the calibration range.
Keywords:Low concentration measurement  spectrophotometry  uranium-thiocyanate system  analysis of variance  errors in measurement
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏