首页 | 本学科首页   官方微博 | 高级检索  
     检索      


HPLC analysis of Cr, V and Mo using pre- in combination with on-column derivatisation by oxine, bipyridine and H2O2
Authors:Yi-he Wu and Georg Schwedt
Institution:(1) Shanghai Institute of Mechanical Engineering, 516 Jun Gong Lu, 200093 Shanghai, People's Republic of China;(2) Institute of Food and Analytical Chemistry, University of Stuttgart, Pfaffenwaldring 55, D-7000 Stuttgart 80, Federal Republic of Germany
Abstract:Summary The HPLC behaviours of Cr(VI), Mo(VI) and V(V) peroxo complexes in a H2O2-8-hydroxyquinolinebipyridine system were studied by using pre-column in combination with on-column derivatisation. The chromatograms of Cr(VI), Mo(VI) and V(V) show them to be CrO2– 4, oxine-Mo peroxo and oxine-V-bipyridine peroxo complexes, respectively, and were used for the separation, identification and determination of Cr(VI), Mo(VI) and V(V) using acetonitrile-water as mobile phase. The calibration curves obtained for 20 mgrl injections were linear for 1.4–7.0 mg/l Cr, 1.3–6.5 mg/l Mo and 0.7–3.4 mg/l V. The relative standard deviations were between 6 and 10%.
HPLC-Analyse von Cr, V und Mo unter Verwendung von Vorsäulen- in Kombination mit Säulenderivatisierung durch Oxin, Bipyridin und H2O2
Keywords:
本文献已被 SpringerLink 等数据库收录!