首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Blind source separation for ambulatory sleep recording.
Authors:Fabienne Porée  Amar Kachenoura  Hervé Gauvrit  Catherine Morvan  Guy Carrault  Lotfi Senhadji
Institution:INSERM, U642, Laboratoire Traitement du Signal et de l'Image, Université de Rennes 1, 35042 Rennes, France. Fabienne.Poree@univ-rennes1.fr
Abstract:This paper deals with the conception of a new system for sleep staging in ambulatory conditions. Sleep recording is performed by means of five electrodes: two temporal, two frontal and a reference. This configuration enables to avoid the chin area to enhance the quality of the muscular signal and the hair region for patient convenience. The electroencephalopgram (EEG), eletromyogram (EMG), and electrooculogram (EOG) signals are separated using the Independent Component Analysis approach. The system is compared to a standard sleep analysis system using polysomnographic recordings of 14 patients. The overall concordance of 67.2% is achieved between the two systems. Based on the validation results and the computational efficiency we recommend the clinical use of the proposed system in a commercial sleep analysis platform.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号