Entwicklung und Eichung einer DTA-Apparatur für Temperaturen bis 1500°C |
| |
Authors: | U Strobel O Henning |
| |
Institution: | 1. Sektion Baustoffverfahrenstechnik der Hochschule für Architektur und Bauwesen, Weimar, D.D.R
|
| |
Abstract: | A laboratory-built DTA-apparatus was calibrated by means of enthalpies of fusion (I), lattice transitions (II), endothermic (III) and exothermic reactions (IV). The temperature-dependence of the factor of calibration A is characterized for reactions of types I and II up to 1500°, and for those of types III and IV up to 600°, by an equation of the formA=155.9+ 10?7·T 3. The error of the enthalpies for reactions of types I, II and IV is 10%. The error can be larger for reactions of type III. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|