首页 | 本学科首页   官方微博 | 高级检索  
     检索      


ETAAS determination of nickel in serum and urine
Authors:Todorovska Nadica  Karadjova Irina  Stafilov Trajce
Institution:Institute of Preventive Medical Care and Toxicology, Military Health Institution Center, 1000 Skopje, Macedonia.
Abstract:Methods for the direct determination of Ni in human blood serum and urine by electrothermal atomic absorption spectrometry (ETAAS) are described. Hydrogen peroxide was proposed as matrix modifier, assisting thermal decomposition of proteins during the ashing step. A pyrolysis temperature of 1,200 degrees C was found to be optimal while 2,100 degrees C and 2,200 degrees C were found to be optimal atomizing temperatures for Ni in serum and urine respectively. Calibration was performed by using a calibration curve prepared with aqueous standard solutions of Ni (glycine must be used as modifier for Ni in aqueous solutions). The limits of detection, defined as the blank values plus 3 times the standard deviation of the blank values, were 0.2 microg/L for both serum and urine samples. Relative standard deviations for serum samples with concentrations of Ni in the range 0.5-2 microg/L were 10-15% and for urine samples with Ni concentrations in the range 0.5-2.5 microg/L were 8-10%.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号