首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Air temperature stabilization in the thermally isolated optical laboratory
Authors:J Budagov  V Glagolev  M Lyablin  G Shirkov  H Mainaud Durand
Institution:1. JINR, Dubna, Russia
2. CERN, Meyrin, Switzerland
Abstract:For the studies and calibration of optoelectronic components of the high-precision laser based metrology systems the large volume (50 m3) thermoisolated lab based on seismic isolated concrete block is created. The inside lab volume temperature stabilization for the daily observation at 16.5°C is ±0.05°C with ±0.015°C temperature difference between maximal space separated points. This work was initiated by the needs of high-precision alignment of accelerator components of the CLIC, ILC-type colliders.
Keywords:
本文献已被 SpringerLink 等数据库收录!