Lead ultra-trace on-line preconcentration and determination using selective solid phase extraction and electrothermal atomic absorption spectrometry: applications in seawaters and biological samples
Authors:E Vereda Alonso  M T Siles Cordero  A García de Torres  J M Cano Pavón
Institution:(1) Department of Analytical Chemistry Faculty of Sciences, University of Málaga, 29071 Málaga, Spain
Abstract:In this work, a new chelating resin 1,5-bis (2-pyridyl)-3-sulphophenyl methylene] thiocarbonohydrazide immobilised on aminopropyl-controlled pore glass (550 A; PSTH-cpg) was synthesised and packed in a microcolumn which replaced the sample tip of the autosampler arm. The system was applied to the preconcentration of lead. When microliters of 10% HNO3, which acts as elution agent, pass through the microcolumn, the preconcentrated Pb(II) is eluted and directly deposited in a tungsten-rhodium coated graphite tube. With the use of the separation and preconcentration step and the permanent modifiers, the analytical characteristics of the technique were improved. The proposed method has a linear calibration range from 0.012 to 10 ng ml(-1) of lead. At a sample frequency of 36 h(-1) with a 90 s preconcentration time, the enrichment factor was 20.5, the detection and determination limits were 0.012 and 0.14 ng ml(-1), respectively and the precision, expressed as relative standard deviation, was 3.2% (at 1 ng ml(-1)). Results from the determination of Pb in biological certified reference materials were in agreement with the certified values. Seawaters and other biological samples were analysed too.
Keywords:Electrothermal atomic absorption spectrometry  On-line preconcentration  Lead  Permanent chemical modified  Seawaters  Biological samples
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号