首页 | 本学科首页   官方微博 | 高级检索  
     


Contribution to the determination of the sensitivity of an ionization transducer
Authors:V. A. Moskalev  G. I. Sergeev  V. V. Shashov
Affiliation:(1) Scientific-Research Institute of Electron Introscopy, USSR;(2) S. M. Kirov Polytechnic Institute, Tomsk
Abstract:An analytic expression for the coefficient of secondary-electron collection has been obtained for an ionization transducer operating at pressures of 0.5–3 MPa. The coefficient is shown to be limited in the given case by the space charge of the beam being measured. A probability relation has been constructed for the ionization of air by electrons of various energies under normal conditions. A method is described for the calibration of the ionization transducer. The probability of the ionization of the residual gas at the pressures indicated has been determined experimentally.Translated from Izvestiya Vysshikh Uchebnykh Zavedenii, Fizika, No. 3, pp. 18–21, March, 1985.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏