Trace analysis of gallium arsenide by graphite furnace AAS
Authors:E Beinrohr  Gergely and J Izák
Institution:(1) Department of Analytical Chemistry, Slovak Technical University, Radlinského 9, CS-812 37 Bratislava, Czechoslovakia
Abstract:Summary Crystalline and powdered gallium arsenide samples are analysed by graphite furnace AAS. The sample is dissolved in hydrochloric acid and bromine, ammonium chloride as matrix modifier is added and the matrix elements are selectively evaporated as chlorides and bromides in the graphite tube of the atomizer before the atomization step. The temperature programs are optimized. The accuracy of the results was checked by recovery tests and by ICP-OES. The detection limits (mgrg/g) are 0.1 (Ca, Fe), 0.05 (Zn, Pb), 0.03 (Mg), 0.01 (Al, Ni, Cu), 0.005 (Co), 0.002 (Cr) and 0.001 (Mn).
Spurenanalyse von Galliumarsenid durch Graphitrohr-AAS
Keywords:
本文献已被 SpringerLink 等数据库收录!