Facility for non-destructive analysis for major and trace elements using neutron-capture gamma-ray spectrometry |
| |
Authors: | D L Anderson M P Failey W H Zoller W B Walters G E Gordon R M Lindstrom |
| |
Institution: | (1) Department of Cehmistry, University of Maryland, 20742 College Park, Maryland, (USA);(2) Center for Analytical Chemistry, National Bureau of Standards, 20234 Washington, D.C., (USA);(3) Present address: Babcock and Wilcox, LRC, 24505 Lynchburg, Virginia |
| |
Abstract: | A facility for neutron-capture γ-ray spectroscopy for analytical purposes has been developed and tested at the National Bureau
of Standards reactor. The system consists of an internal beam tube with collimators, an external beam tube and irradiation
station, a Compton-suppressed Ge(Li) γ-ray detection system, and a minicomputer-based data-collection and-analysis system.
Detection limits have been established for many elements and errors arising from neutron self shielding, γ-ray peak overlap,
neutron beam variations, and sample matrix evaluated. |
| |
Keywords: | |