首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Facility for non-destructive analysis for major and trace elements using neutron-capture gamma-ray spectrometry
Authors:D L Anderson  M P Failey  W H Zoller  W B Walters  G E Gordon  R M Lindstrom
Institution:(1) Department of Cehmistry, University of Maryland, 20742 College Park, Maryland, (USA);(2) Center for Analytical Chemistry, National Bureau of Standards, 20234 Washington, D.C., (USA);(3) Present address: Babcock and Wilcox, LRC, 24505 Lynchburg, Virginia
Abstract:A facility for neutron-capture γ-ray spectroscopy for analytical purposes has been developed and tested at the National Bureau of Standards reactor. The system consists of an internal beam tube with collimators, an external beam tube and irradiation station, a Compton-suppressed Ge(Li) γ-ray detection system, and a minicomputer-based data-collection and-analysis system. Detection limits have been established for many elements and errors arising from neutron self shielding, γ-ray peak overlap, neutron beam variations, and sample matrix evaluated.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号