Method of Measuring the Spectral Shape of the Radiation Line of a Three-Micron Long-Path Laser Gas Analyzer |
| |
Authors: | S V Ivanov V Ya Panchenko |
| |
Institution: | (1) Institute of the Problems of Laser and Information Technologies, Russian Academy of Sciences, 2 Pionerskaya Str., Troitsk, Moscow Region, 142190, Russia |
| |
Abstract: | A new method to experimentally determine the spectral shape of the radiation line of a tunable long-path absorption laser gas analyzer operating in a three-micron range ( = 2.5–4.5 m) on the basis of a difference-frequency generator is suggested. The method is based on measurement of the transmission function of a gas medium upon modulation of its optical thickness or pressure. Specific vibrational–rotational transitions and also optimal pressures of CO2, N2O, and CH4 are recommended for carrying out experiments. |
| |
Keywords: | laser gas analyzer radiation line shape measurement method |
本文献已被 SpringerLink 等数据库收录! |