首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Comparison of an instrumental and modified Kjeldahl technique for determination of nitrogen in niobium and tantalum alloys
Authors:Merkle E J  Graab J W  Davis W F
Institution:

Lewis Research Center, National Aeronautics and Space Administration, Cleveland Ohio, U.S.A.

Abstract:Results obtained for the determination of nitrogen in two tantalum alloys and six niobium alloys by modified Kjeldahl and Leco TC-30 nitrogen—oxygen determinator are compared. In the 5–25 ppm range, for tantalum alloys, the relative standard deviation was 3–9% by the Kjeldahl procedure and 9–11% by the instrumental technique. In the range 30–80 ppm, for niobium alloys, the relative standard deviation was 2–8% by the Kjeldahl procedure and 5–7% by the instrumental technique.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!