Ultrathin fiber probes with extended depth of focus for optical coherence tomography |
| |
Authors: | Lorenser Dirk Yang Xiaojie Sampson David D |
| |
Affiliation: | Optical+Biomedical Engineering Laboratory, School of Electrical, Electronic & Computer Engineering, The University of Western Australia, WA 6009, Australia. dirk.lorenser@uwa.edu.au |
| |
Abstract: | We report on a novel scheme for extending the depth of focus (DOF) of ultrathin (125 μm diameter) fiber probes for optical coherence tomography (OCT) using a simple phase mask consisting of graded-index (GRIN) fiber. The technique is compatible with existing all-in-fiber probe fabrication techniques, and our simulations show that it can provide a DOF gain of ~2 at a modest ~5 dB reduction of peak sensitivity. In a prototype device using commercially available GRIN fiber, a DOF gain of 1.55 is obtained, validated by beam profiling and OCT imaging. |
| |