首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of total mercury in environmental and biological samples by flow injection cold vapour atomic absorption spectrometry
Authors:James Murphy  Phil Jones  Steve J Hill
Institution:

Department of Environmental Science, University of Plymouth, Drake Circus, Plymouth, PL4 8AA, UK

Abstract:A simple and accurate method has been developed for the determination of total mercury in environmental and biological samples. The method utilises an off-line microwave digestion stage followed by analysis using a flow injection system with detection by cold vapour atomic absorption spectrometry.

The method has been validated using two certified reference materials (DORM-1 dogfish and MESS-2 estuarine sediment) and the results agreed well with the certified values. A detection limit of 0.2 ng g?1 Hg was obtained and no significant interference was observed. The method was finally applied to the determination of mercury in river sediments and canned tuna fish, and gave results in the range 0.1–3.0 mg kg?1.

Keywords:Cold vapour atomic absorption spectrometry  Flow Injection  Mercury  Microwave digestion
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏