Simulations of Redox Mediation within Bioactive Hydrogels of Amperometric Biosensors |
| |
Authors: | Walter Torres |
| |
Affiliation: | Department of Chemical and Biomolecular Engineering , Center for Bioelectronics, Biosensors and Biochips, Clemson University , Clemson, SC |
| |
Abstract: | Highly hydrated bioactive hydrogels containing immobilized oxidoreductase enzymes and immobilized redox mediators were simulated as the biorecognition layer of amperometric biosensors. The linear dynamic range of the amperometric response of mediated biosensors increases and moves to higher concentration brackets with an increase in the concentration of mediator. This informs the design of biosensors that target the same analyte but possesses several independently addressable electrodes modified with hydrogels that contain different concentrations of mediator. Increases in enzyme concentration increase the linear dynamic range but does not alter the sensitivity of amperometric biosensors. Both sensitivity and linear dynamic range of mediated amperometric enzyme biosensors may be “tuned” by varying the concentrations of the enzyme and the mediator. Simulations effectively guide the initial domains of study of complex systems such as implantable biosensors. |
| |
Keywords: | simulation enzymes hydrogels mediation biosensors |
|
|