首页 | 本学科首页   官方微博 | 高级检索  
     


Preatomization behaviour of Se(-II), Se(IV), and Se(VI) in the graphite furnace without and with matrix modifiers
Authors:V. Krivan and M. Kückenwaitz
Affiliation:(1) Sektion Analytik und Höchstreinigung, Universität Ulm, Albert-Einstein-Allee 11, W-7900 Ulm, Federal Republic of Germany
Abstract:Summary Using 75Se as a radiotracer, the preatomization behaviour of selenium in the graphite furnace was studied. The selenium forms investigated included Se(-II)-methionine, selenite, and selenate in a 0.2% HNO3 solution, and in a 0.2% HNO3 solution containing 1% NaCl. The effect of nickel nitrate and of the mixture of palladium/magnesium nitrates as matrix modifiers and of boron nitride coating of the graphite tube on the behaviour of selenium was investigated. The best stabilization effect for all oxidation states of selenium in the conventional graphite tube was achieved by using the mixture Pd/Mg. A considerable degree of modifier-free stabilization of selenuium could be achieved in boron nitride coated tubes. After the conversion of Se(IV) to a volatile piaselenol, a quantitative preatomization separation of Se(IV) from Se(VI) in the boron nitride coated tube was possible. However problems with these newtype tubes still to be solved include the need to increase the thermal stability of the coating.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏