Instrumentation for simultaneous multielement atomic absorption spectrometry with graphite furnace atomization
Authors:James M. Harnly
Affiliation:(1) Agricultural Research Service, Beltsville Human Nutrition Research Center, Food Composition Laboratory, US Department of Agriculture, Building 161, BARC-East, Beltsville, MD 20705, USA
Abstract:Graphite furnace-atomic absorption spectrometry (GF-AAS) is restricted to the determination of 4 to 6 elements simultaneously due to the limitations of hollow cathode lamps. However, a consideration of prototype continuum source instruments and recent advances in the fields of spectrometer and detector technology suggests that a multielement GF-AAS instrument, with the multielement versatility associated with atomic emission spectrometry, is possible. Such a multielement instrument would employ a continuum source and provide 1.) multielement determinations for 30 to 40 elements, 2.) wavelength and time integrated absorbance measurements which are independent of the source width, 3.) detection limits comparable to line source AAS with the potential for another order of magnitude improvement using atomization at elevated pressures, 4.) extended calibration ranges limited only by the memory of the atomizer, and 5.) high resolution inspection of the spectra surrounding the analytical wavelength. Such an instrument could provide figures of merit comparable to inductively coupled plasma-mass spectrometer with considerably less complexity.