首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development of a compact supersonic jet/multiphoton ionization/time-of-flight mass spectrometer for the on-site analysis of dioxin, part I: Evaluation of basic performance.
Authors:Junichi Matsumoto  Bunji Nakano  Totaro Imasaka
Institution:Department of Applied Chemistry, Graduate School of Engineering, Kyushu University, Hakozaki, Higashi-ku, Fukuoka 812-8581, Japan.
Abstract:A new type of compact supersonic jet/resonance-enhanced multiphoton ionization/time-of-flight mass spectrometer is described. The analytical instrument, consisting of a single turbo molecular pump equipped with a rotary pump, was maintained at < 2 x 10(-3) Pa when a 0.3-atm sample was injected into a vacuum at 10-Hz using a 200-micros pulse valve. The diameters of the extraction and ground skimmer electrodes were expanded to 30 mm in order to avoid strong focusing and defocusing of the ion, and the optimum conditions for the system were investigated. The mass spectrometer functioned as expected: (1) no defocusing of the ion beam was observed even when the potential of the einzel lens was adjusted to zero; (2) the direction of the ion beam to an assembly of microchannel plates deviated in the expected manner when the potential of the defection electrode was changed from 0 to 30 V.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏