Amperometric Determination of Creatinine with a Dialysis Membrane‐Covered Nitrobenzene/Water Interface for Urine Analysis
Authors:Risa Matsui  Toru Sakaki  Toshiyuki Osakai
Institution:1. Department of Chemistry, Graduate School of Science, Kobe University, Nada, Kobe 657‐8501, Japan;2. A&T Corporation, 2023‐1 Endo, Fujisawa, Kanagawa 252‐0816, Japan
Abstract:The ion transfer of creatinine (CRE) at a polarized nitrobenzene (NB)/water (W) interface has been studied. When the pH of the W phase is in the range of 1.2 to 4.0, a well‐defined voltammetric wave is observed for a simple transfer of CRE+ (protonated form) at the NB/W interface. This transfer reaction has been applied to develop an amperometric method for the determination of CRE in urine. In this system, the NB/W interface is covered with a dialysis membrane to prevent possible interference from urine proteins. The concentration of CRE in a urine control has successfully been determined.
Keywords:Creatinine  Amperometry  ITIES  Dialysis membrane  Urine analysis