首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Dual-jet plasmatron for medical applications
Authors:I V Osipova  I A Ryabkov  N G Vnukova  N V Bulina  G N Churilov
Institution:(1) L. V. Kirenskii Institute of Physics, Siberian Branch, Russian Academy of Sciences, 50 Akademgorodok, Krasnoyarsk, 660036;(2) N. S. Karpovich Emergency Hospital, Krasnoyarsk, Russia
Abstract:We present a design for a dual-jet arc plasmatron operating at a frequency of 66 kHz in an argon flow at atmospheric pressure. We present the results of determination of the temperature, electron concentration, and electrode erosion obtained by atomic emission spectral analysis. The proposed convenient design for a dualjet plasmatron and the low erosion of the copper electrodes in the plasma make it possible to use it for medical purposes. Report given at the Fifth International Conference on Plasma Physics and Plasma Technologies (PPPT-5), 18–22 September 2006, Minsk, Belarus. __________ Translated from Zhurnal Prikladnoi Spektroskopii, Vol. 74, No. 1, pp. 139–140, January–February, 2007.
Keywords:dual-jet plasmatron  arc plasma  electrode erosion
本文献已被 SpringerLink 等数据库收录!