Simulation study on active noise control for a 4-T MRI scanner |
| |
Authors: | Li Mingfeng Lim Teik C Lee Jing-Huei |
| |
Affiliation: | Department of Mechanical, Industrial and Nuclear Engineering, University of Cincinnati, Cincinnati, OH 45221, USA. |
| |
Abstract: | The purpose of this work is to study computationally the possibility of the application of a hybrid active noise control technique for magnetic resonance imaging (MRI) acoustic noise reduction. A hybrid control system combined with both feedforward and feedback loops embedded is proposed for potential application on active MRI noise reduction. A set of computational simulation studies were performed. Sets of MRI acoustic noise emissions measured at the patient's left ear location were recorded and used in the simulation study. By comparing three different control systems, namely, the feedback, the feedforward and the hybrid control, our results revealed that the hybrid control system is the most effective. The hybrid control system achieved approximately a 20-dB reduction at the principal frequency component. We concluded that the proposed hybrid active control scheme could have a potential application for MRI scanner noise reduction. |
| |
Keywords: | MRI Acoustic noise Active noise control |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|