首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of selenium in food matrices by replicate sample neutron activation analysis
Authors:M. G. Ventura  M. C. Freitas  A. M. G. Pacheco
Affiliation:(1) Reactor-ITN, Technological and Nuclear Institute, E.N. 10, 2686-953 Sacavém, Portugal;(2) REQUIMTE-FCT, New University of Lisbon, Monte da Caparica, 2829-516 Caparica, Portugal;(3) CERENA-IST, Technical University of Lisbon, Av. Rovisco Pais 1, 1049-001 Lisboa, Portugal
Abstract:The replicate sample instrumental neutron activation method was optimized and used for the determination of selenium in foodstuffs. The method was reliable, yielding accurate results. Lower detections limits were obtained after each successive irradiation. Different irradiation conditions were used depending on the type of sample. For samples with higher selenium contents (meat, fish, eggs), the measured selenium in the first replicate is in all cases larger than the detection limit, but a better accuracy was obtained with a larger number of replicates (2–3 replicates). For samples with extremely low selenium contents (vegetable samples), at least seven replicates were necessary to obtain a concentration value two times larger than the detection limit.
Keywords:Selenium  Food  Portugal  RSINAA
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏