Accurate quantitative analysis of gold alloys using multi-pulse laser induced breakdown spectroscopy and a correlation-based calibration method
Authors:G  bor Galb  cs, Nikoletta Jedlinszki, G  bor Cseh, Zolt  n Galb  cs,L  szl    ri
Affiliation:

aDepartment of Inorganic and Analytical Chemistry, University of Szeged, 6720 Szeged, Dóm tér 7., Hungary

Abstract:Multi-pulse laser induced breakdown spectroscopy (LIBS), in combination with the generalized linear correlation calibration method (GLCM), was applied to the quantitative analysis (fineness determination) of quaternary gold alloys. Accuracy and precision on the order of a few thousandths (‰) was achieved. The analytical performance is directly comparable to that of the standard cupellation method (fire assay), but provides results within minutes and is virtually non-destructive, as it consumes only a few micrograms of the sample.
Keywords:Double-pulse   Gold assay   Laser induced plasma spectroscopy   LIBS   Linear correlation
本文献已被 ScienceDirect 等数据库收录!