Fully automated determination of N-nitrosamines in environmental waters by headspace solid-phase microextraction followed by GC-MS-MS |
| |
Authors: | Llop Anna Borrull Francesc Pocurull Eva |
| |
Institution: | Department of Analytical Chemistry and Organic Chemistry, Universitat Rovira i Virgili, Spain. |
| |
Abstract: | A fully automated method for determining nine Environmental Protection Agency N-nitrosamines in several types of environmental waters at ng/L levels is presented. The method is based on a headspace solid-phase microextraction followed by GC-MS-MS using chemical ionization. Three different fibers (carboxen/PDMS, divinylbenzene/carboxen/PDMS, and PEG) were tested. Solid-phase microextraction conditions were best when a divinylbenzene/carboxen/PDMS fiber was exposed for 60?min in the headspace of 10?mL water samples at pH 7 containing 360?g/L of NaCl, at 45°C. All compounds were analyzed by GC-MS-MS within 18?min. The method was validated using effluent from an urban wastewater treatment plant and the LODs ranged from 1 to 5?ng/L. The method was then applied to determine the N-nitrosamines in samples of different complexities, such as tap water and several influent and effluent wastewater samples from urban and industrial wastewater treatment plants and a potable water treatment plant. Although the analysis of influent industrial wastewater revealed high concentrations of some compounds (N-nitrosomorpholine and N-nitrosodimethylamine at μg/L levels), in industrial effluents and other samples, the concentrations were substantially lower (ng/L levels). The new method is suitable for the simple and reliable determination of N-nitrosamines in highly complex water samples in a completely automated procedure. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|