首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Effect of Input Waveform to Vibration Speaker on Sound Quality of Electric Artificial Voice
Authors:T Asakura  K Shindo
Institution:Tokyo University of Science, Chiba, 278-8510, Japan
Abstract:A number of studies have been conducted on the improvement of the sound quality of electrical artificial laryngeal speech, the speech produced has been difficult to hear compared to a natural voice. For this reason, it is necessary to effectively improve the frequency characteristics of the input signal. In the present study, to improve the sound quality of vocalization using an electrical artifi- cial larynx, first, a comparison of the frequency characteristics between the real and artificial voices was conducted, and three filters that can make the frequency characteristics of the artificial voice closer to those of a real voice were generated. Then, the influence of these filters on the quality of the artificial voice was investigated via physical measurement and a subjective evaluation experiment targeted at Japanese five vowels. It was found that the intelligibility of artificial /a/ and /o/ sounds was improved, whereas little improvement was observed in the case of /i/, /u/, and /e/. The obtained results confirmed the effect of optimizing the input signal into the vibration speaker and indicated areas for further improvement.
Keywords:Electriolaryngeal voice  intelligibility  naturalness  loudness
点击此处可从《声与振动》浏览原始摘要信息
点击此处可从《声与振动》下载免费的PDF全文