首页 | 本学科首页   官方微博 | 高级检索  
     检索      


High-performance liquid chromatographic determination of boric acid and germanic acid with conductometric detection
Authors:Tetsuo Okada and Tooru Kuwamoto
Institution:(1) Department of Chemistry, Faculty of Science, Kyoto University, Kyoto, Japan;(2) Present address: Faculty of Liberal Arts, Shizuoka University, 422 Shizuoka, Japan
Abstract:Summary Boric and germanic acids were detected by a conductivity detector after ion-exclusion Chromatographic separation on a cation-exchange resin. Neither acid could be directly measured by a conductivity detector because they were very weak. However, use of eluents of polyol compounds permitted their determination by a conductivity detector. With 0.1 M of fructose as eluent, it was possible to detect 0.5 ppb of boric acid (as B) and 1.5 ppb of germanic acid (as Ge).
HPLC-Bestimmung von Borsäure und Germaniumsäure mit konduktometrischer Detection
Zusammenfassung Beide Säuren wurden mit Hilfe eines Leitfähigkeitsdetektors nach Trennung durch Ionenausschluß-Chromatographie an einem Kationenaustauscher analysiert. Eine direkte Messung war wegen der Schwäche der Säuren nicht möglich; daher wurde eine Elution mit Polyolen durchgeführt. Mit 0,1 M Fructoselösung als Eluens konnten noch 0,5 ppb Borsäure (als B) und 1,5 ppb Germaniumsäure (als Ge) nachgewiesen werden.
Keywords:
本文献已被 SpringerLink 等数据库收录!